A failing Samsung Medison C2-5ET does not always mean buying a new probe. Use the fault table below to match what you are seeing on screen to a likely cause, then check what a repair assessment would need.
| Manufacturer | Samsung Medison |
|---|---|
| Model | C2-5ET |
| Probe type | Convex array |
| Compatible systems | Medison SonoAce Pico, Medison SonoAce SA-6000II, Medison SonoAce SA-8000, Medison SonoAce X4 |
You will usually see the C2-5ET running on Medison SonoAce Pico, Medison SonoAce SA-6000II, Medison SonoAce SA-8000, Medison SonoAce X4. Mention the console model with the fault description – connector and revision differences matter.
Curved arrays like the C2-5ET see heavy abdominal duty, and the curved lens face takes the brunt of it – delamination and membrane wear typically show up before element failures do.
Common failure modes for the C2-5ET
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Latch or locking hardware damaged on the connector shell | Broken locking mechanics on the connector housing | Usually repairable |
| Depth performance clearly below what this probe used to deliver | Ageing elements or lens delamination reducing acoustic output | Case by case |
| Cable jacket cracked, wrinkled or discolored | Surface jacket wear – the conductors inside are often still good | Usually repairable |
| Intermittent signal that changes as the cable moves | Broken internal conductors in the cable | Case by case |
| Strain relief separated, split or stiff | Flex fatigue where the cable meets the probe or connector | Usually repairable |
| Vertical shadowing or missing scan lines on screen | Element dropout, or a conductor break inside the cable | Case by case |
| Large areas of the array not firing | Widespread element or acoustic stack failure | Usually not economical |
Check these before you ship the probe
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Flex the cable gently along its length while imaging – intermittent dropout points to broken conductors.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Run the system’s built-in probe test or element check if your console provides one.
Repair, exchange or replace — which applies to your C2-5ET
Three paths apply to a failing C2-5ET: repair (typical for lens, housing, cable and connector damage), exchange (fastest route back to scanning) or replacement (when the acoustic stack is too far gone). The deciding factors are how much of the array still fires, whether the housing is intact and what downtime costs you. Each probe is judged on its actual condition.
What we need to quote your C2-5ET repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific C2-5ET can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.






Reviews
There are no reviews yet.