A transducer fails, a clinician flags it, and now the question lands on your desk: ultrasound probe repair or replace? Biomedical engineers and imaging managers make this call under pressure, usually with incomplete information and a scanner sitting idle. This page is the hub of our fault library. It gives you a working framework — what drives the decision, which fault types are usually repairable, and where the economics flip — with links to the detailed page for each specific failure.
What it looks like
The decision moment arrives in one of two forms. Either the probe has visible physical damage — a drop, a crushed cable, a torn strain relief — or the image has quietly degraded until someone refuses to scan with it. On screen, the classic complaints are dropout lines, artifacts, and signal noise; a dark band that tracks a dead element group; or a general loss of penetration and contrast that no preset change fixes. We cover the on-screen family in detail under element dropout and degraded image quality.
The physical family is easier to spot: delamination, holes, and physical wear on the probe face (see lens delamination and lens wear and scratches), cracked housings (housing damage), cracked strain reliefs (strain relief repair), cable whitening or kinks (cable damage), and bent connector pins (connector repair). Specialty probes have their own failure signatures: bite marks and articulation problems on TEE probes (TEE probe repair guide), leak-test failures on endocavity transducers (fluid ingress and leak testing), and stalled or grinding sweeps on mechanical volume probes (3D/4D mechanical repair).
Why it happens
Almost every probe failure traces back to one of four mechanisms. Mechanical trauma: drops, cart collisions, cable run-overs, TEE bite damage. Wear: lens abrasion from thousands of scans, cable flex fatigue at the strain relief, articulation cables stretching out of spec. Fluid ingress: a breached lens or housing seam lets gel or disinfectant reach the array and the wiring, which is why leak testing matters so much on endovaginal and endorectal transducers. Then there’s acoustic aging — crystals and their bonding layers degrade over years of service, showing up as gradual sensitivity loss rather than a single event.
The mechanism matters because it predicts scope. Trauma tends to be localized: one crushed cable section, one torn boot, one cracked lens edge. Aging and fluid ingress tend to be widespread: many elements down, corrosion across the array interconnect. Localized damage is where repair earns its keep. Widespread damage is where the economics start working against you.
Check it yourself before shipping anything
Ten minutes of testing on site sharpens the decision considerably.
- Swap test. Move the probe to another port, then to another console if you have one. If the fault follows the probe, it’s the probe. If it stays with the port, stop — you have a console problem, not a transducer problem.
- Uniformity check. Scan air with high gain and uniform TGC. Count the dark vertical bands and note their width. One or two narrow lines reads very differently from a third of the image gone.
- Cable flex test. With a live image, gently flex the cable along its length and at both strain reliefs. Flickering or dropout that tracks your hand points to a cable or termination fault, which changes the repair scope.
- Physical inspection. Look at the lens under good light for cuts, bubbles, or peeling. Check the housing seams. Look into the connector for bent pins before you seat it again — forcing a damaged connector into a console can turn a probe repair into a console repair.
- For endocavity and TEE probes: if you have a leak tester, run it. A failed leak test changes the urgency; that probe should come out of clinical rotation immediately.
Write down what you find. Those notes become the fault description on your quote request, and they shorten the assessment on our end.
Is it repairable?
There’s no single answer for “a broken probe,” which is exactly why this framework exists. The honest pattern across fault types looks like this.
Usually repairable: localized mechanical damage. Lens replacement, housing reseal, strain relief tear repair, cable retermination or replacement, connector pin and shell work, TEE tip replacements and articulation cable replacement, insertion tube restoration, and mechanical drive repairs on 3D/4D probes. These faults live in the parts of the probe that were built to be serviced or replaced around a healthy array.
Case by case: element dropout and image-quality faults. A few dead elements at the array edge is a different problem from a dead cluster mid-array. Crystal dropout and interference can stem from the cable, the connector, or the array itself — the location decides the repair path and the cost. Fluid ingress caught early, before corrosion spreads, also sits here.
Usually not economical: widespread element loss, arrays killed by long-term fluid ingress, and acoustic aging across the aperture. Array replacements exist as a repair category, but on many models the array is most of the probe’s value. When the core imaging component is gone, you’re not repairing a probe — you’re rebuilding one, and the numbers rarely justify it.
The dividing line, in one sentence: repair works when the array survives and the damage sits around it.
Repair or replace
Four factors decide this, and they interact. Weigh them together, not one at a time.
| Factor | Leans repair | Leans replace |
|---|---|---|
| Damage scope | Localized: lens, housing, cable, connector, strain relief, mechanics | Widespread: large element loss, corroded array, aged acoustics |
| Array condition | Uniformity test clean or near-clean | Multiple wide dropout bands, dead zones mid-array |
| Probe value and availability | Current premium probes — TEE, 3D/4D, specialty transducers | Cheap, plentiful standard probes for older consoles |
| Downtime tolerance | Exchange or loaner can bridge the gap, or a spare exists | No spare, no exchange stock, clinical schedule can’t wait |
A few practical notes on applying it. High-value probes shift the math hard toward repair: a TEE or 3D/4D transducer carries enough replacement cost that even substantial repair work can make sense. Commodity linear probes for a console you plan to retire shift the other way. Would you repair a probe for a scanner leaving service next year? Probably not, unless the repair is minor.
Downtime deserves its own line because it’s the factor managers underweight until it bites. A repairable probe you can’t spare is still a problem. This is where exchange programs earn their place in the framework: a tested replacement probe ships out while your damaged unit becomes trade-in stock, collapsing the downtime question entirely. When we assess your probe, the quote can include an exchange option alongside the repair path where stock allows, so you’re comparing real numbers, not guesses.
Fleet context matters too. If you run a mixed fleet of GE, Philips, and Mindray systems, the same fault can carry different economics per platform — probe pricing and parts availability vary widely between manufacturers. Portable fleets built on SonoSite hardware have their own patterns, since those probes absorb more transport abuse. Older Siemens Acuson and Toshiba probes often lean toward repair simply because clean replacements are getting scarce. Our brand hubs and model reference pages cover the specifics per platform.
What a quote needs
Four things let us give you a real answer instead of a range.
- Label photo. The probe’s ID label with full model and part number. This is how we distinguish variants that look identical but differ internally.
- Fault description. What the operator sees, plus your self-check results — swap test outcome, dropout band count, whether flexing the cable changes anything, leak test result if applicable.
- Console model. The system the probe runs on. It matters for testing and for confirming compatibility if exchange enters the picture.
- Damage photos. Clear shots of the lens face, housing seams, strain reliefs, cable, and connector pins. For image faults, a screen capture of the uniformity test is worth a paragraph of description.
Send everything through our contact page. Every probe gets a physical assessment before any work starts, and the findings are confirmed in writing — what failed, whether it’s repairable, and what the repair or exchange path costs — so the repair-or-replace decision is yours to make with real information in hand.
Frequently asked questions
The probe still images but has one thin dropout line. Should I keep using it or send it in now?
A single narrow band means a small, localized element or interconnect fault, and catching it early usually keeps the repair scope small. You can often keep scanning short-term if the band sits outside the clinical region of interest, but document it, monitor whether it widens, and get an assessment quoted. Dropout that spreads or is joined by noise suggests a progressing fault such as fluid ingress, and that probe should come out of rotation.
How does an exchange probe differ from a repair, and when does it make sense?
Repair returns your own probe after the fault is corrected; exchange ships you a tested equivalent probe while your damaged unit is traded in. Exchange makes sense when downtime is the binding constraint — no spare probe, a full clinical schedule — or when your unit's damage is widespread enough that repair is usually not economical but the probe still has trade-in value. Where stock allows, we quote both paths side by side so you can compare.
Can you tell me over email whether my probe is repairable before I ship it?
We can give a preliminary read from a label photo, damage photos, and a good fault description — localized damage like a torn strain relief or cracked lens photographs well and is usually repairable. What photos can't show is array condition underneath, which is why the final verdict comes from a physical assessment on our test equipment, confirmed in writing before any work begins. If assessment shows repair isn't economical, you'll know before committing.
Indexed Probe Repair & Biopsy Guide Directory — Group 6 (501 Models)
SIEMENS Transducer & Guide Models (102)
Siemens (Acuson) P10-4 · Siemens (Acuson) P5-1 · Siemens (Acuson) P8-3 · Siemens (Acuson) P9-4 · Siemens (Acuson) PA023E · Siemens (Acuson) PH4-1 · Siemens (Acuson) PL13 · Siemens (Acuson) PL28 · Siemens (Acuson) PX4-1 · Siemens (Acuson) S219 · Siemens (Acuson) S228 · Siemens (Acuson) S319 · Siemens (Acuson) S3194 · Siemens (Acuson) S328 · Siemens (Acuson) S510B · Siemens (Acuson) S519 · Siemens (Acuson) S5192 · Siemens (Acuson) S519C · Siemens (Acuson) S519R · Siemens (Acuson) S7146 · Siemens (Acuson) S714C · Siemens (Acuson) S714R · Siemens (Acuson) TE-V7B · Siemens (Acuson) TE-V7M · Siemens (Acuson) V2 · Siemens (Acuson) V219 · Siemens (Acuson) V2C · Siemens (Acuson) V319 · Siemens (Acuson) V328 · Siemens (Acuson) V366 · Siemens (Acuson) V4 · Siemens (Acuson) V4c · Siemens (Acuson) V4R · Siemens (Acuson) V5 · Siemens (Acuson) V510B · Siemens (Acuson) V519C · Siemens (Acuson) V519R · Siemens (Acuson) V5C · Siemens (Acuson) V5Ms · Siemens (Acuson) V5R · Siemens (Acuson) V7 · Siemens (Acuson) V705B · Siemens (Acuson) V714C · Siemens (Acuson) V714R · Siemens (Acuson) V714S · Siemens (Acuson) V714T · Siemens (Acuson) V7C · Siemens (Acuson) V7R · Siemens (Acuson) VF12-4 · Siemens (Acuson) VF13-5SP · Siemens (Acuson) VF16-5 · Siemens (Acuson) VF7-3 · Siemens (Acuson) VFX 13-5 · Siemens (Acuson) VFX9-4 · Siemens (Acuson) Z6Ms · Siemens C5-2-FREESTYLE · Siemens C6-2 · Siemens CA123 · Siemens CA431 · Siemens CW5-CW · Siemens EC1123 · Siemens (Acuson) EC9-4 · Siemens EC9-4 · Siemens EC9-4-S · Siemens EC9-4-X · Siemens ENDO-V-PRIMA · Siemens ENDO-V-SI · Siemens ENDO-V-SONOLINE · Siemens EV9-4 · Siemens EV9F4 · Siemens IOE323 · Siemens L10-5 · Siemens L13-5-FREESTYLE · Siemens L17-5-FREESTYLE · Siemens L8-3-FREESTYLE · Siemens LA522E · Siemens P4-2 · Siemens P8-4 · Siemens PA122e · Siemens (Acuson) MC9-4 · Siemens (Acuson) EC10C5 · Siemens EV-8C4 · Siemens (Acuson) EC9-4(X300) · Siemens (Acuson) 9EVF4 · Siemens (Acuson) 4V1 · Siemens (Acuson) C5 · Siemens (Acuson) 14L5 · Siemens (Acuson) 9L4 · Siemens (Acuson) VF13-5 · Siemens (Acuson) 18L6HD · Siemens (Acuson) 6C1HD · Siemens (Acuson) CH6-2 · Siemens (Acuson) EC9-4(S2000) · Siemens V5M · Siemens VF10-5 · Siemens VF13-5-ANTARES · Siemens VF13-5 · Siemens VF13-5-SONOLINE · Siemens VF13-5-X · Siemens VF8-3 · Siemens VFX13-5-ANTARES · Siemens VFX13-5-ELEGRA
SIUI Transducer & Guide Models (20)
SIUI 4DL40K · SIUI C3IC · SIUI C3L60K · SIUI C3LC · SIUI C5IC · SIUI C5LF · SIUI C6LC · SIUI ECBP · SIUI L10LC · SIUI L8LC · SIUI P3FC · SIUI P5FC · SIUI V6L11C · SIUI U5LC · SIUI U7I50 · SIUI V6L11A · SIUI V6l11c · SIUI V6L11G · SIUI V6L11K · SIUI V6LC
SONOSCAPE Transducer & Guide Models (51)
Sonoscape 10I2 · Sonoscape 10L1 · Sonoscape 10L2 · Sonoscape 2P1 · Sonoscape 2P2 · Sonoscape 5P1 · Sonoscape 5P2 · Sonoscape 5V1 · Sonoscape 6V1 · Sonoscape 6V3 · Sonoscape 6V4 · Sonoscape 6V7 · Sonoscape 7P-B · Sonoscape 8P1 · Sonoscape C312 · Sonoscape C344 · Sonoscape C351 · Sonoscape C354 · Sonoscape C542 · Sonoscape C543 · Sonoscape C611 · Sonoscape C612 · Sonoscape EC2 · Sonoscape EC9-5 · Sonoscape L741 · Sonoscape L742 · Sonoscape L745 · Sonoscape L746 · Sonoscape L752 (S8) · Sonoscape L752 (S/P Series) · Sonoscape L773 · Sonoscape 6V1 · Sonoscape 6V3 · Sonoscape 6V4 · Sonoscape 6V7 · Sonoscape C322 · Sonoscape 7P-A · Sonoscape C612 · Sonoscape C613 · Sonoscape S1-5 · Sonoscape 10L-I · Sonoscape 4P-A · Sonoscape C1-6 · Sonoscape L742 · Sonoscape C344 · Sonoscape 3C-A · Sonoscape 12L-A · Sonoscape C353 · Sonoscape L741 · Sonoscape C351 · Sonoscape VC6-2
SONOSITE Transducer & Guide Models (86)
Sonosite 8-3 · Sonosite 8Ce · Sonosite C11 · Sonosite C11e · Sonosite C11n · Sonosite C11x · Sonosite C11xp · Sonosite C15 · Sonosite C15e · Sonosite C35X · Sonosite C35Xp · Sonosite C60 · Sonosite C60E · Sonosite C60n · Sonosite C60v · Sonosite C60X · Sonosite C60xi · Sonosite C60Xp · Sonosite C8 · Sonosite C8x · Sonosite D2 Pencil/CW · Sonosite D2x Pencil/CW · Sonosite D2xp Pencil/CW · Sonosite HFL38 · Sonosite HFL38e · Sonosite HFL38x · Sonosite HFL38Xi · Sonosite HFL38xp · Sonosite HFL50x · Sonosite HFL50xp · Sonosite HSL25x · Sonosite HSL25xp · Sonosite HST · Sonosite IC10-3 · Sonosite ICT · Sonosite ICT7-4 · Sonosite ICT8-5 · Sonosite ICTx · Sonosite ICTxp · Sonosite L12-3 · Sonosite L15-4 · Sonosite L19-5 · Sonosite L25 · Sonosite L25e · Sonosite L25n · Sonosite L25v · Sonosite L25x · Sonosite L25xp · Sonosite L38 · Sonosite L38e · Sonosite L38n · Sonosite L38v · Sonosite L38x · Sonosite L38Xi · Sonosite L38Xi · Sonosite L38xp · Sonosite L52 · Sonosite L52x · Sonosite ICT · Sonosite P21X · Sonosite L25 · Sonosite HFL38X · Sonosite L38xi · Sonosite C60 · Sonosite P10 · Sonosite P10x · Sonosite P10xp · Sonosite P11x · Sonosite P17 · Sonosite P21n · Sonosite P21v · Sonosite P21X-5-1 · Sonosite P21x · Sonosite P21xp · Sonosite P5-1 · Sonosite RC60XI · Sonosite rP19x · Sonosite RP19xp · Sonosite SLA/13-6 · Sonosite SLAx · Sonosite SLT · Sonosite SLT/10-5 · Sonosite TEE/8-3 · Sonosite TEEx · Sonosite TEExi · Sonosite TEEXp
SUMMIT Transducer & Guide Models (3)
TERASON Transducer & Guide Models (3)
THREE Transducer & Guide Models (1)
TOSHIBA Transducer & Guide Models (191)
Toshiba Canon PSI-30VX · Toshiba CANON-PVT-375BT · Toshiba CANON-PVT-745BTF · Toshiba IOE-703H · Toshiba PEF-508SC · Toshiba PEF-510MA · Toshiba PEF-510MB · Toshiba PEF-510SB · Toshiba PEF-704LA · Toshiba PEK-510MB · Toshiba PEM-508SC · Toshiba PET-508M · Toshiba PET-508MA · Toshiba PET-510MA · Toshiba PET-510MB · Toshiba PET-511BTM · Toshiba PET-512MC · Toshiba PET-805LA (12LI4) · Toshiba PLE-308M · Toshiba PLE-705S · Toshiba PLF-308P · Toshiba PLF-503NT · Toshiba PLF-703NT · Toshiba PLF-703ST · Toshiba PLF-705S · Toshiba PLF-805ST · Toshiba PLG-308P · Toshiba PLG-506M · Toshiba PLG-805S · Toshiba PLI-1205BX (i18LX5) · Toshiba PLI-2002BT (i22LH8) · Toshiba PLI-2004BX (i24LX8) · Toshiba PLI-705BX (i11LX3) · Toshiba PLK-703AT · Toshiba PLM-1202S · Toshiba PLM-1204AT · Toshiba PLM-1204S · Toshiba PLM-503AT · Toshiba PLM-703AT · Toshiba PLM-805A · Toshiba PLM-805AT · Toshiba PLN-503AT · Toshiba PLN-703AT · Toshiba PLN-805AT · Toshiba PLQ-1203A · Toshiba PLQ-703A · Toshiba PLQ-805A · Toshiba PLT-1005BT · Toshiba PLT-1202BT (17LH7) · Toshiba PLT-1202S · Toshiba PLT-1204AT · Toshiba PLT-1204AT · Toshiba PLT-1204AX · Toshiba PLT-1204BT · Toshiba PLT-1204BX · Toshiba PLT-1204MV · Toshiba PLT-308P · Toshiba PLT-604AT · Toshiba PLT-704AT · Toshiba PLT-704BT · Toshiba PLT-704SBT · Toshiba PLT-705BT · Toshiba PLT-705BTH · Toshiba PLT-805AT · Toshiba PLU-1005BT (14L5) · Toshiba PLU-1202BT (17LH7) · Toshiba PLU-1204BT (18L7) · Toshiba PLU-704BT (11L4) · Toshiba PSF-25FT · Toshiba PSF-25KT · Toshiba PSF-25LT · Toshiba PSF-37CT · Toshiba PSF-37FT · Toshiba PSF-37HT · Toshiba PSF-50AT · Toshiba PSF-50ET · Toshiba PSF-50FT · Toshiba PSH-70LT · Toshiba PSI-30BX · Toshiba PSI-40VX · Toshiba PSK-20CT · Toshiba PSK-25AT · Toshiba PSK-37AT · Toshiba PSK-37CT · Toshiba PSK-50LT · Toshiba PSK-70LT · Toshiba PSM-20CT · Toshiba PSM-25AT · Toshiba PSM-30BT · Toshiba PSM-37AT · Toshiba PSM-37CT · Toshiba PSM-50AT · Toshiba PSM-70AT · Toshiba PSN-25AT · Toshiba PSN-37CT · Toshiba PST-20CT · Toshiba PST-25AT · Toshiba PST-25BT (5S1) · Toshiba PST-25ST · Toshiba PST-30BT · Toshiba PST-37CT · Toshiba PST-50AT · Toshiba PST-50BT · Toshiba PST-65AT · Toshiba PSU-30BT · Toshiba PSU-50BT · Toshiba PVE-375M · Toshiba PVF-357MT · Toshiba PVF-375MT · Toshiba PVF-381MT · Toshiba PVF-575MT · Toshiba PVF-620ST · Toshiba PVF-621VT · Toshiba PVF-641VT · Toshiba PVF-738F · Toshiba PVF-738H · Toshiba PVF-745V · Toshiba PVG-366M · Toshiba PVG-601V · Toshiba PVG-681S · Toshiba PVG-720S · Toshiba PVI-475BT · Toshiba PVI-475BX · Toshiba PVK-357AT · Toshiba PVL-715RT · Toshiba PVL-725RT · Toshiba PVM-375A · Toshiba PVM-375AT · Toshiba PVM-381AT · Toshiba PVM-620ST · Toshiba PVM-621VT · Toshiba PVM-651VT · Toshiba PVM-662AT · Toshiba PVM-672AT · Toshiba PVM-740RT · Toshiba PVM-787LA · Toshiba PVN-375AT · Toshiba PVN-661VT · Toshiba PVN-760ST · Toshiba PVQ-375A · Toshiba PVQ-381A · Toshiba PVQ-641V · Toshiba PVQ-662A · Toshiba PVQ-674A · Toshiba PVT-350BTP · Toshiba PVT-375AT · Toshiba PVT-375BT-A · Toshiba PVT-375BT · Toshiba PVT-375BT · Toshiba PVT-375MV · Toshiba PVT-375SC · Toshiba PVT-375ST · Toshiba PVT-382BT · Toshiba PVT-382BT · Toshiba PVT-382MV · Toshiba PVT-475BT · Toshiba PVT-574BT · Toshiba PVT-575MV · Toshiba PVT-661VT · Toshiba PVT-674BT · Toshiba PVT-674BT · Toshiba PVT-675MV · Toshiba PVT-681MV · Toshiba PVT-712BT · Toshiba PVT-712BT · Toshiba PVT-745BTV · Toshiba PVT-770RT · Toshiba PVT-781VT · Toshiba PVT-781VT · Toshiba PVT-781VT · Toshiba PVT-781VTE · Toshiba PVT-781VTE · Toshiba PVU-375BT-6C1 · Toshiba PVU-375BT · Toshiba PVU-674BT · Toshiba PVU-770ST · Toshiba PVU-781VT · Toshiba PVT-661VT · Toshiba PVT-781VT · Toshiba PVT-681MV · Toshiba PVM-381AT · Toshiba PVT-382BT · Toshiba PST-30BT · Toshiba PSM-25AT · Toshiba PLM-703AT · Toshiba PLT-1204BT · Toshiba PLT-704BT · Toshiba PLG-805S · Toshiba PVT-375BT · Toshiba PVM-375AT · Toshiba PLT-805AT
TWO Transducer & Guide Models (1)
ULTRASONIX Transducer & Guide Models (8)
· Ultrasonix C5-2 60 · Ultrasonix C5-2/60GPS · · Ultrasonix L14-5 · L14-5 · C5-2 · EC9-5
UNIMAGING Transducer & Guide Models (6)
Unimaging L12-5 · Unimaging L14-6 · Unimaging P4-2 · Unimaging L14-6 · Unimaging L12-5 · Unimaging L12-5
VINNO Transducer & Guide Models (2)
WELLD Transducer & Guide Models (2)
WISONIC Transducer & Guide Models (9)
Wisonic C5-1 · Wisonic EV10-4 · Wisonic P4-1 · Wisonic S4-2 · Wisonic L10-5 · Wisonic L15-4B · Wisonic L15-4 · Wisonic C5-1 · Wisonic C5-1B
ZONARE Transducer & Guide Models (12)
Zonare C4-1 · Zonare C8-3 · Zonare C9-3 · Zonare E9-4 · Zonare L14-5W 60 · Zonare L14-5W · Zonare L8-3 · Zonare P4-1C · Zonare E9-4 · Zonare C4-1 · Zonare C9-3 · Zonare L14-5W
Compatible Ultrasound Consoles & Brand Hubs (132)
Siemens Sonoline G60S · Siemens Sonoline Omnia · Siemens Sonoline Prima · Siemens Sonoline SI-200 · Siemens Sonoline SI-250 · Siemens Sonoline SI-400 · Siemens Sonoline SI-450 · Siemens Sonoline Sienna · Siemens Sonoline SL-1 · Siemens Sonoline SL-2 · Siemens Sonoline Versa · Siemens Sonoline Versa Plus · Siemens Sonoline Versa Pro · Siemens/Acuson Cypress · Siemens/Acuson Cypress Plus · SIUI Apogee 1100 · SIUI Apogee 1100V · SIUI Apogee 3800 · SIUI Apogee 5300 · SIUI CTS 485 · SIUI CTS-5000 · Sonoscape A6 · Sonoscape E1 · Sonoscape E2 · Sonoscape E3 · Sonoscape P50 · Sonoscape S2 · Sonoscape S20 · Sonoscape S22 · Sonoscape S40 · Sonoscape S6 · Sonoscape S8 · Sonoscape S8 Exp · Sonoscape S8 Expert · Sonoscape S9 · Sonoscape S9 Pro · Sonoscape SSI-1000 · Sonoscape SSI-5000 · Sonoscape SSI-600 · Sonoscape SSI-6000 · Sonoscape SSI-800 · SonoSite 180 Plus · SonoSite Edge · Sonosite Edge II · Sonosite iViz · SonoSite M-Turbo · SonoSite MicroMaxx · SonoSite MicroMaxx Vet · Sonosite PX · Sonosite SII · SonoSite Titan · Sonosite Titan Vet · Sonosite X-Porte · SSD-3500/4000 · SSD3500/3500plus/4000 · Summit LifeDop L150 · Summit LifeDop L250 · Summit LifeDop L350 · Toshiba Aplio 300 (TUS-A300) · Toshiba Aplio 400 (TUS-A400) · Toshiba Aplio 50 (SSA-700A) · Toshiba Aplio 50 (SSA-750A) · Toshiba Aplio 500 (TUS-A500) · Toshiba Aplio 80 (SSA-770A) · Toshiba Aplio MX (SSA-780A) · Toshiba Aplio XG (SSA-790A) · Toshiba Artida (SSH-880CV) · Toshiba Capasee (SSA-220) · Toshiba Capasee (SSA-220A) · Toshiba CoreVision (SSA-350A) · Toshiba Eccocee (SSA-340A) · Toshiba Famio 5 (SSA-510A) · Toshiba Famio 8 (SSA-530A) · Toshiba JustVision 200 (SSA-320A) · Toshiba JustVision 400 (SSA-325A) · Toshiba Nemio 10 (SSA-550A-10) · Toshiba Nemio 17 (SSA-550A-17) · Toshiba Nemio 17 (SSA-550A) · Toshiba Nemio 20 (SSA-550A-20) · Toshiba Nemio 30 (SSA-550A-30) · Toshiba Nemio 35 · Toshiba Nemio XG (SSA-580A) · Toshiba PowerPace (SSH-140A) · Toshiba PowerVision 6000 (SSA-370A) · Toshiba PowerVision 7000 (SSA-380A) · Toshiba PowerVision 8000 (SSA-390A) · Toshiba Sonolayer (SSA-250A) · Toshiba Sonolayer (SSA-270A) · Toshiba Sonolayer (SSH-160A) · Toshiba Tosbee (SSA-240A) · Toshiba Viamo (SSA-640A) · Toshiba Xario (SSA-660A) · Toshiba Xario XG (SSA-680A) · Ultrasonix SonixMDP · Ultrasonix SonixOP · Ultrasonix SonixSP · Ultrasonix SonixTouch · Vingmed System Five · Vivid 3 · Vivid 7 · Vivid E · Vivid E9 · Vivid E95 · Vivid i · Vivid IQ · Vivid q · Vivid S5 · Vivid S6 · Vivid S70 · Vivid series · Vivid T8 · Voluson 730 · Voluson e · Voluson E10 · Voluson E6 · Voluson E8 · Voluson E8 Expert · Voluson E9 · Voluson i · VOLUSON P8 · Voluson portable · Voluson S6 · Voluson S8 · Voluson series · Voluson women · WS80 · WS80A · Guía de Reparación · Guia de Reparo · Learning Centre · Probe Repair Guide · 未分类
All Clinical Engineering & Diagnostic Guides (76)
How to Repair the Ultrasound Probe? · Where Can I Repair the Ultrasound Probe? · Repair Fukuda FUT · Repair Aloka UST · Repair GE C36 Probe: Ultimate Guide to C… · Who repairs Alpinion L3 · Where can BK 8670 ultrasound probe be re… · Who can repair BK 8819/E10C4/9019 ultras… · When you need to repair Konica L18 · Ultrasound Probe Element Dropout: Diagno… · Degraded Ultrasound Image Quality: Probe · Ultrasound Probe Lens Delamination: Bubb… · Ultrasound Probe Lens Scratches and Memb… · Cracked or Dented Ultrasound Probe Housi… · Ultrasound Probe Strain Relief Damage an… · Ultrasound Probe Cable Damage: Jacket We… · Ultrasound Probe Connector Damage: Pins,… · TEE Probe Repair: Bite Damage, Articulat… · Endocavity Probe Fluid Ingress and Leak … · 3D/4D Probe Mechanical Failures: Motor a… · Repair or Replace an Ultrasound Probe: A… · Reparo de Transdutor TEE: Mordida, Artic… · Dano no Cabo do Transdutor de Ultrassom:… · Dano no Conector do Transdutor de Ultras… · Falha de Elementos em Transdutor de Ultr… · Carcaça de Transdutor de Ultrassom Trinc… · Qualidade de Imagem de Ultrassom Degrada… · Delaminação da Lente do Transdutor de Ul… · Riscos na Lente e Desgaste da Membrana d… · Reparar ou Trocar um Transdutor de Ultra… · Dano e Reparo no Alívio de Tensão do Tra… · Falhas Mecânicas em Transdutores 3D/4D: … · Infiltração de Fluido e Teste de Vazamen… · Fallas Mecánicas en Transductores 3D/4D:… · Infiltración de Fluido en el Transductor… · Reparación de Transductor TEE: Daño por … · Daño del Cable del Transductor de Ultras… · Daño del Conector del Transductor de Ult… · Falla de Elementos en Transductor de Ult… · Carcasa del Transductor de Ultrasonido A… · Calidad de Imagen de Ultrasonido Degrada… · Delaminación de la Lente del Transductor… · Rayones en la Lente y Desgaste de la Mem… · Reparar o Reemplazar un Transductor de U… · Daño y Reparación del Protector de Cable… · Probe Fault or System Fault? A Three · Before You Ship an Ultrasound Probe for … · How to Evaluate a Used or Refurbished Ul… · A One · Transdutor de Ultrassom: Tipos, Estrutur… · Manutenção Preventiva de Transdutores de… · Transdutor Endocavitário: Tipos, Cuidado… · Transdutor Linear: Aplicações, Frequênci… · Transdutor Convexo: Uso, Frequências e F… · Transdutor Setorial (Phased Array): Uso … · Transdutor Convexo GE: Modelos, Compatib… · Transdutor Mindray: Modelos, Compatibili… · Transdutor de Ultrassom Philips: Modelos… · Calibração de Ultrassom: O Que É, O Que … · Falha do Transdutor ou do Aparelho? Árvo… · Antes de Enviar um Transdutor para Repar… · Como Avaliar um Transdutor Usado ou Reco… · Rotina Mensal de Verificação de Transdut… · Transdutor Siemens: Modelos, Compatibili… · Transdutor Esaote: Modelos, Compatibilid… · Transdutor Toshiba: Modelos, Compatibili… · Ultrasound Probe Model Cross · Ultrasound Probe Lens & Crystal Arti… · Tabela Cruzada de Modelos de Transdutor … · Tabla Cruzada de Modelos de Transductor … · Autoavaliação de Artefatos e Falhas em T… · Autoevaluación de Artefactos y Fallas en… · Ultrasound Probe Not Recognized & Re… · Transdutor de Ultrassom Não Reconhecido … · Ultrasound Biopsy Needle Guides: Reusabl… · Transductor de Ultrasonido No Reconocido…