Ultrasound Probe Cable Damage: Jacket Wear to Conductor Breaks

A cracked or whitened cable is the fault biomedical engineers most often ignore until the image starts flickering. If your probe cable shows jacket cracks, deep wrinkles, or discoloration — or the image drops out when someone moves the cable — this page tells you whether ultrasound probe cable repair will fix it, or whether you’re looking at a full cable replacement. The short version: outer jacket damage is usually repairable; broken internal conductors are case by case. The flex test below tells you which one you have.

Connector Cable Strain relief Housing Array stack Lens
Where this fault sits on the probe — highlighted section
Cable Fault Symptom Electrical / Mechanical Root Cause Bench Diagnostic & Repair Path
Horizontal snow flashes or RF interference when moving cable Fractured braided coaxial EMI shield or loose shield ground termination inside handle or connector. Locate break point via TDR / flex test; re-terminate shield ground or replace main cable assembly.
Intermittent black dropout lines that appear/disappear on flex Fatigued 40–46 AWG micro-coax center conductors breaking contact under tension. Micro-coax conductor splicing (if localized near boot) or complete multi-coax harness replacement.
Crushed cable from cart wheel roll-over Short-circuit between adjacent coax channels or pinched transmit lines causing console fault code. Isolate shorted lines on continuity/capacitance analyzer and replace damaged cable harness assembly.

What it looks like

Technician working on an ultrasound probe cable at an extraction-fitted bench
Cable and strain-relief work is done under extraction, with the probe fully supported.

Cable damage shows up in two places: on the cable itself, and on the screen.

On the cable, look for:

  • Cracked or split jacket — usually near the strain relief at the probe end or connector end, where the cable bends hardest. Sometimes the crack runs lengthwise along a section that’s been pinched in a cart drawer.
  • Cable whitening or discoloration — the gray jacket turns chalky white, yellow, or blotchy. This is not cosmetic. It’s the polymer breaking down, and the jacket in that zone will crack next.
  • Badly wrinkled or kinked sections — the cable holds a sharp bend even when you straighten it, or feels lumpy where the internal bundle has shifted.
  • Soft or spongy spots — a sign the shield braid has separated or gel has migrated under the jacket.

On the screen, the classic sign is intermittent image on flex: the image is fine until the sonographer repositions, then you get flickering, a burst of noise, dropped lines, or a complete signal loss that comes back when the cable settles. Broken conductors can also mimic element dropout — dark vertical bands in the image — because each coax in the bundle carries specific channels. If you’re seeing persistent dark bands with no cable movement involved, read the element dropout page instead; the flex test below separates the two.

Why it happens

A probe cable is a bundle of dozens to hundreds of micro-coaxial conductors, each thinner than a hair, wrapped in shielding and a flexible jacket. Every part of that construction has a distinct failure mode.

Mechanical fatigue. The cable flexes thousands of times a week. The highest stress concentrates in the first few centimeters past the strain relief, which is why most jacket cracks and most conductor breaks cluster there. Cart wheels rolling over the cable, drawers pinching it, and probes left dangling by the cable all add localized crush damage — the same category of abuse that produces bite marks, punctures, and crushed internal components on handheld probes.

Chemistry damage. Cable whitening is almost always chemical, not mechanical. Disinfectant wipes, gel residue, and cleaning agents that are fine on the probe body slowly leach plasticizers out of the cable jacket. The polymer loses flexibility, turns white or chalky, and then cracks under normal bending. If your whole department’s cables are whitening at the same rate, the cause is your wipe-down protocol, not one careless user. Check the disinfectant against the probe manufacturer’s compatibility list — GE, Philips, and Mindray each publish their own, and they don’t agree with each other.

Conductor breaks. Once the jacket is compromised, or after enough flex cycles even with an intact jacket, individual micro-coax conductors fracture. A single broken coax kills or degrades the channels it carries. A fatigued section breaks progressively — one conductor this month, five more over the next quarter. That’s why an intermittent flex fault left alone becomes a permanent multi-channel fault.

Shield degradation. A cracked jacket also lets the shield braid corrode or separate, which shows up as increased noise and interference across the whole image rather than discrete dead channels.

Check it yourself before shipping anything

Ten minutes of bench checks will tell you most of what a repair lab’s first-day assessment finds.

1. The flex test. This is the diagnostic. Connect the probe, get a live image (a phantom or your own forearm is fine), and have a second person watch the screen while you work along the cable. Grip the cable every 10–15 cm and flex it gently side to side. Do the same at both strain reliefs. If flexing a specific section makes the image flicker, drop lines, or fill with noise — and it recovers when you release — you’ve localized a conductor break to that section. Note where it is. A fault that only appears at the probe-end strain relief is a different repair than one in the mid-cable run, and that distinction drives the quote.

2. Rule out the connector. Wiggle the cable right at the connector housing while watching the image, then reseat the connector in a different port on the console. If the fault follows the port or responds to pressure on the connector body rather than the cable, you may have a connector-side fault — see the connector repair page.

3. Inspect the full length under good light. Run the cable through your fingers end to end. Mark every crack, whitened zone, kink, and soft spot with tape. Photograph each one with something for scale.

4. Check the strain reliefs. A torn or hardened strain relief is often the root cause of the cable damage next to it, and repairing the cable without addressing the strain relief invites a repeat failure. Cracked strain reliefs are covered on their own page.

5. Swap-test if you can. Same probe model on a different console, or a known-good probe on the suspect console. Two minutes, and it stops you from shipping a probe when the actual fault is a console channel board.

Is it repairable?

Split the verdict by what’s actually broken.

Jacket damage with intact conductors — usually repairable. If the flex test is clean and the image is stable, cracks, whitening, and wrinkling are outer-layer problems. Jacket section repair and resealing restores the biocompatible, sealed outer surface before fluid reaches the shield and conductors. Catch it at this stage and the repair is straightforward. This applies across probe families — a whitened jacket on a Toshiba convex probe and on a SonoSite point-of-care probe is the same chemistry and the same fix.

Internal conductor breaks — case by case. The deciding factor is location. Breaks concentrated near either end can often be handled by cable retermination: cutting back past the fatigued section and re-landing the conductors at the array interconnect or the connector. Breaks in the middle of the run, or fatigue spread along the whole length, mean the cable itself is done, and the question becomes whether a full cable replacement is available and economical for that model. On high-channel-count or specialty probes, cable replacement is a significant job; on some models it’s routine.

Widespread whitening with brittleness — usually not economical. When chemistry has degraded the jacket along most of its length, patching one section leaves nine more waiting to crack. That probe usually needs full cable replacement or an honest repair-or-replace conversation.

One warning: an intermittent flex fault does not stabilize. It progresses. The probe that flickers occasionally today is the probe with permanent dead channels next quarter — and by then the repair scope has grown.

Repair or replace

For cable faults specifically, the decision logic runs like this:

  • Jacket-only damage: repair, almost regardless of probe value. It’s preventive — you’re stopping a cheap fault from becoming an expensive one.
  • End-zone conductor breaks on a mid-to-high-value probe: get a retermination quote. On standard linear, convex, and phased array probes from major platforms this is well-trodden work.
  • Mid-cable breaks or full-length fatigue: the quote will reflect a cable replacement, so weigh it against the replacement cost of the probe itself. On a common workhorse model with a healthy used market, replacement may win. On specialty probes — TEE, 3D/4D, high-frequency — the probe’s replacement cost usually keeps repair on the table.
  • Cable damage plus other faults (lens wear, housing cracks, dead elements): tally the combined scope before deciding. Stacked repairs on one probe change the math. The repair-or-replace decision guide walks through the full calculation.

Whatever you decide, don’t run a probe with exposed shield or conductors on patients. A compromised cable jacket is an electrical safety and infection-control issue, not just an image quality one.

What a quote needs

A cable fault quote is only as accurate as what you send. Four things:

  • A photo of the probe label — model and serial, legible. This tells the lab the exact cable construction and whether retermination or replacement parts exist for that model.
  • Your flex test result — where along the cable the fault triggers, or “flex test clean, jacket damage only.” This single line often decides retermination versus replacement before the probe ships.
  • Console make and model — the system you tested on, so the assessment can be run on matching or compatible hardware.
  • Photos of every damage point — each crack, whitened zone, and kink you marked during inspection, with scale.

Every repair starts with a bench assessment, and the findings and verdict are confirmed in writing before any work proceeds — so you’re deciding on evidence, not a guess. Send the details through the contact page and include the flex test result; it’s the one piece of information most quotes arrive without.

Frequently asked questions

The image only flickers when the sonographer moves the cable. Can I keep using the probe until it gets worse?

You can, but you're trading repair scope for time. An intermittent flex fault means one or more micro-coax conductors have already fractured and are making contact only in certain cable positions. Flex fatigue is progressive: the same bending that broke the first conductor keeps working on its neighbors. A fault that today needs retermination of one end can grow into multi-section damage that forces full cable replacement. If the jacket is intact and it's purely intermittent, schedule the repair; if there's any exposed shield or conductor, pull the probe from patient use immediately.

Our probe cables are turning white but the images are still fine. Is that a problem?

Yes, and it's a chemistry problem, not wear. Whitening means disinfectants or gel residue are leaching plasticizers out of the jacket polymer. The whitened zone becomes brittle and will crack under normal flexing, which then exposes the shield and conductors to fluid. If multiple probes in the department are whitening together, audit your wipe-down products against the probe manufacturer's compatibility list before you repair anything, or the replacement jackets will whiten the same way. Localized whitening with a clean flex test is usually repairable at the jacket level.

How do I tell a cable conductor break from element dropout in the array?

Use the flex test. Get a live image and flex the cable section by section while watching the screen. If dark bands or noise appear and disappear as you flex a specific spot, the fault is in the cable. If dark vertical bands sit in the same image position no matter what you do to the cable, the fault is more likely dead elements or array-side damage. The two can coexist on an abused probe, which is why a written bench assessment checks both before quoting.

Indexed Probe Repair & Biopsy Guide Directory — Group 4 (502 Models)

LEAPMED Transducer & Guide Models (215)

Leapmed 120026 Reusable · Leapmed 120027 Reusable · Leapmed 120029 Reusable · Leapmed 120031 Reusable · Leapmed 120033 Reusable · Leapmed 120037 Reusable · Leapmed 120038 Reusable · Leapmed 120039 Reusable · Leapmed 120040 Reusable · Leapmed 120042 Reusable · Leapmed 120043 Reusable · Leapmed 120044 Reusable · Leapmed 120045 Reusable · Leapmed 120046 Reusable · Leapmed 120047 Reusable · Leapmed 120048 Reusable · Leapmed 120049 Reusable · Leapmed 120050 Reusable · Leapmed 120051 Reusable · Leapmed 120052 Reusable · Leapmed 120053 Reusable · Leapmed 120054 Reusable · Leapmed 120055 Reusable · Leapmed 120056 Reusable · Leapmed 120057 Reusable · Leapmed 120058 Reusable · Leapmed 120059 Reusable · Leapmed 120060 Reusable · Leapmed 120061 Reusable · Leapmed 120062 Reusable · Leapmed 120063 Reusable · Leapmed 120064 Reusable · Leapmed 120065 Reusable · Leapmed 120066 Reusable · Leapmed 120067 Reusable · Leapmed 120068 Reusable · Leapmed 120069 Reusable · Leapmed 120070 Reusable · Leapmed 120071 Reusable · Leapmed 120072 Reusable · Leapmed 120073 Reusable · Leapmed 120074 Reusable · Leapmed 120075 Reusable · Leapmed 120076 Reusable · Leapmed 120078 Reusable · Leapmed 120079 Reusable · Leapmed 120080 Reusable · Leapmed 120081 Reusable · Leapmed 120082 Reusable · Leapmed 120083 Reusable · Leapmed 120084 Reusable · Leapmed 120085 Reusable · Leapmed 120086 Reusable · Leapmed 120087 Reusable · Leapmed 120088 Reusable · Leapmed 120089 Reusable · Leapmed 120090 Reusable · Leapmed 120091 Reusable · Leapmed 120092 Reusable · Leapmed 120093 Reusable · Leapmed 120094 Reusable · Leapmed 120095 Reusable · Leapmed 120096 Reusable · Leapmed 120097 Reusable · Leapmed 120098 Reusable · Leapmed 120099 Reusable · Leapmed 120100 Reusable · Leapmed 120101 Reusable · Leapmed 120102 Reusable · Leapmed 120103 Reusable · Leapmed 120104 Reusable · Leapmed 120105 Reusable · Leapmed 120106 Reusable · Leapmed 120107 Reusable · Leapmed 120108 Reusable · Leapmed 120109 Reusable · Leapmed 120111 Reusable · Leapmed 120112 Reusable · Leapmed 120113 Reusable · Leapmed 120114 Reusable · Leapmed 120115 Reusable · Leapmed 120116 Reusable · Leapmed 120117 Reusable · Leapmed 120119 Reusable · Leapmed 120120 Reusable · Leapmed 120121 Reusable · Leapmed 120122 Reusable · Leapmed 120123 Reusable · Leapmed 120124 Reusable · Leapmed 120125 Reusable · Leapmed 120126 Reusable · Leapmed 120127 Reusable · Leapmed 120128 Reusable · Leapmed 120129 Reusable · Leapmed 120130 Reusable · Leapmed 120131 Reusable · Leapmed 120132 Reusable · Leapmed 120133 Reusable · Leapmed 120134 Reusable · Leapmed 120135 Reusable · Leapmed 120136 Reusable · Leapmed 120137 Reusable · Leapmed 120138 Reusable · Leapmed 120139 Reusable · Leapmed 120140 Reusable · Leapmed 120141 Reusable · Leapmed 120142 Reusable · Leapmed 120143 Reusable · Leapmed 120144 Reusable · Leapmed 120145 Reusable · Leapmed 120146 Reusable · Leapmed 120147 Reusable · Leapmed 120149 Reusable · Leapmed 120150 Reusable · Leapmed 120151 Reusable · Leapmed 120152 Reusable · Leapmed 120153 Reusable · Leapmed 120154 Reusable · Leapmed 120155 Reusable · Leapmed 120156 Reusable · Leapmed 120157 Reusable · Leapmed 120158 Reusable · Leapmed 120159 Reusable · Leapmed 120160 Reusable · Leapmed 120161 Reusable · Leapmed 120162 Reusable · Leapmed 120163 Reusable · Leapmed 120164 Reusable · Leapmed 120165 Reusable · Leapmed 120166 Reusable · Leapmed 120167 Reusable · Leapmed 120168 Reusable · Leapmed 120169 Reusable · Leapmed 120170 Reusable · Leapmed 120171 Reusable · Leapmed 120172 Reusable · Leapmed 120173 Reusable · Leapmed 120174 Reusable · Leapmed 120175 Reusable · Leapmed 120176 Reusable · Leapmed 120177 Reusable · Leapmed 120178 Reusable · Leapmed 120179 Reusable · Leapmed 120180 Reusable · Leapmed 120181 Reusable · Leapmed 120182 Reusable · Leapmed 120183 Reusable · Leapmed 120184 Reusable · Leapmed 120185 Reusable · Leapmed 120186 Reusable · Leapmed 120187 Reusable · Leapmed 120188 Reusable · Leapmed 120189 Reusable · Leapmed 120190 Reusable · Leapmed 120191 Reusable · Leapmed 120192 Reusable · Leapmed 120193 Reusable · Leapmed 120194 Reusable · Leapmed 120195 Reusable · Leapmed 120196 Reusable · Leapmed 120197 Reusable · Leapmed 120198 Reusable · Leapmed 120200 Reusable · Leapmed 120201 Reusable · Leapmed 120202 Reusable · Leapmed 120203 Reusable · Leapmed 120204 Reusable · Leapmed 120205 Reusable · Leapmed 120206 Reusable · Leapmed 120207 Reusable · Leapmed 120208 Reusable · Leapmed 120209 Reusable · Leapmed 120210 Reusable · Leapmed 120211 Reusable · Leapmed 120218 Reusable · Leapmed 120219 Reusable · Leapmed 120220 Reusable · Leapmed 120221 Reusable · Leapmed 120222 Reusable · Leapmed 120223 Reusable · Leapmed 120224 Reusable · Leapmed 120225 Reusable · Leapmed 120226 Reusable · Leapmed 120227 Reusable · Leapmed 120228 Reusable · Leapmed 120229 Reusable · Leapmed 120230 Reusable · Leapmed 120231 Reusable · Leapmed 120232 Reusable · Leapmed 120233 Reusable · Leapmed 120235 Reusable · Leapmed 120236 Reusable · Leapmed 120237 Reusable · Leapmed 120239 Reusable · Leapmed 120240 Reusable · Leapmed 120241 Reusable · Leapmed 120242 Reusable · Leapmed 120243 Reusable · Leapmed 120244 Reusable · Leapmed 120245 Reusable · Leapmed 120246 Reusable · Leapmed 120247 Reusable · Leapmed 120248 Reusable · Leapmed 120249 Reusable · Leapmed 120251 Reusable · Leapmed 120252 Reusable · Leapmed 120253 Reusable · Leapmed 120254 Reusable · Leapmed 120255 Reusable · Leapmed 120277 Reusable · Leapmed 120329 Reusable · Leapmed 120342 Reusable · Leapmed 120345 Reusable · Leapmed 120412 Reusable · Leapmed 121001 Reusable

MEDICAL Transducer & Guide Models (1)

Akicare MEDICAL Support & Replac

MINDRAY Transducer & Guide Models (205)

Mindray 10L24EA · Mindray 10L4 · Mindray 10L4s · Mindray 2P2P · Mindray 2P2s · Mindray 35C20EA · Mindray 35C50EA · Mindray 35C50EB · Mindray 3C1 · Mindray 3C1s · Mindray 3C5A · Mindray 3C5P-35C50P-2 · Mindray 3C5P · Mindray 3C5S · Mindray 4CD4s · Mindray 50L60EAV · Mindray 65C15EA · Mindray 65C15EAV · Mindray 65C15EB · Mindray 65EC10EA EB · Mindray 65EC10EB · Mindray 65EC10ED · Mindray 65EL60EA · Mindray 6C2 · Mindray 6C2a · Mindray 6C2P · Mindray 6C2s · Mindray 6CV1 · Mindray 6CV1 · Mindray 6CV1 V10-4 V10-4B · Mindray 6CV1P · Mindray 6CV1s · Mindray 6LB7 · Mindray 6LB7E · Mindray 6LB7s · Mindray 6LE7P · Mindray 6LE7s · Mindray 75L38EA · Mindray 75L38EB · Mindray 75L50EAV · Mindray 75L53EA · Mindray 75L60EA · Mindray 75LT38EA · Mindray 7L4A · Mindray 7L4B · Mindray 7L4BP · Mindray 7L4P · Mindray 7L4s · Mindray 7L5 · Mindray 7L5P · Mindray 7L5S · Mindray 7L6s · Mindray 7LT4P · Mindray 7LT4P-T · Mindray 7LT4s · Mindray 7LT4S-T · Mindray 7TL4s · Mindray A2-CONTINUOUS-WAVE · Mindray A5-CONTINUOUS-WAVE · Mindray C10-3 · Mindray C11-3e · Mindray C11-3s · Mindray C11-3U · Mindray C18-5 · Mindray C3-E · Mindray C4-1s · Mindray C4-1U · Mindray C5-1E · Mindray C5-1s · Mindray C5-1U · Mindray C5-2 · Mindray C5-2e · Mindray C5-2s · Mindray C6-1 · Mindray C6-2 · Mindray C6-2GE · Mindray C6-2gs · Mindray C6-2GU · Mindray C6-2P · Mindray C7-3e · Mindray C9-3sp · Mindray CB10-4e · Mindray CB10-4P · Mindray CW2s Pencil/CW · Mindray CW5s Pencil/CW · Mindray D6-2 · Mindray D6-2e · Mindray D6-2P · Mindray D7-2 · Mindray D7-2E · Mindray D7-2s · Mindray D8-2U · Mindray DE10-3e · Mindray DE10-3WE · Mindray DE10-3WU · Mindray DE11-3E · Mindray DE11-3s · Mindray E9-3 (3D) · Mindray E9-3 · Mindray ELC13-4s · Mindray Hitachi EUP-ES52E · Mindray 6CV1 · Mindray SC5-1E · Mindray L10-3s · Mindray L11-3U · Mindray L11-3VNs · Mindray L11-4 · Mindray L11-4 · Mindray L12-3e · Mindray L12-3RCs · Mindray L12-4s · Mindray L13-3 · Mindray L13-3N · Mindray L13-3Ns · Mindray L13-3s · Mindray L14-3Ws · Mindray L14-5sp · Mindray L14-5WE · Mindray L14-5WU · Mindray L14-6 · Mindray L14-6N · Mindray L14-6ne · Mindray L14-6ns · Mindray L14-6P · Mindray L14-6s · Mindray L14-6we · Mindray L14-6Ws · Mindray L14-6WU · Mindray L16-4HE · Mindray L16-4Hs · Mindray L16-4HU · Mindray L20-5 · Mindray L20-5E · Mindray L20-5s · Mindray L20-5U · Mindray L7-3 · Mindray L7-3e · Mindray L7-3s · Mindray L9-3E · Mindray L9-3s · Mindray L9-3U · Mindray LM14-6E · Mindray LM14-6s · Mindray LM16-4U · Mindray 35C50EA · Mindray 75L38EA · Mindray 6C2 · Mindray 3C5s · Mindray 7L4 · Mindray 3C1 · Mindray 6LB7 · Mindray 2P2 · Mindray C5-2 · Mindray C7-3E · Mindray P7-3 · Mindray C5-1 · Mindray C6-2G · Mindray L11-3U · Mindray SC5-1U · Mindray P10-4e · Mindray P10-4s · Mindray P10-4U · Mindray P12-4S · Mindray P4-2e · Mindray P4-2NE · Mindray P4-2s · Mindray P7-3c · Mindray P7-3e · Mindray P7-3P · Mindray P7-3s · Mindray P7-3Te · Mindray P7-3ts · Mindray P7-3TU · Mindray P7-3U · Mindray P8-3m · Mindray P8-3TEE Pencil/CW · Mindray P8-3TS · Mindray SC5-1e · Mindray SC5-1N · Mindray SC5-1NS · Mindray SC6-1E · Mindray SC6-1s · Mindray SC6-1U · Mindray SC8-2s · Mindray SC8-2U · Mindray SD8-1E · Mindray SD8-1s · Mindray SP5-1e · Mindray SP5-1N · Mindray SP5-1NS · Mindray SP5-1s · Mindray SP5-1U · Mindray V10-4 · Mindray V10-4b · Mindray V10-4BP · Mindray V10-4s · Mindray V11-3 · Mindray V11-3be · Mindray V11-3e · Mindray V11-3HE · Mindray V11-3Hs · Mindray V11-3HU · Mindray V11-3s · Mindray V11-3we · Mindray V11-3ws

NEWMAN Transducer & Guide Models (2)

Newman D2W · Newman D5

PHILIPS Transducer & Guide Models (79)

Philips 11-3L (21356A) · Philips 15-6L (21390A) · Philips 15-6L · Philips 21255B · Philips 21258B · Philips 21362A · Philips 21365A · Philips 3D6-2 · Philips 3D6 · Philips 3D8-4 · Philips 3D9-3V · Philips 3D9-3V · Philips BP10-5EC · Philips BPT9-5 · Philips BPTRT9-5 · Philips C10-3V · Philips C10-4EC · Philips C10 · Philips C3540 (21321A) · Philips C3540 (21353B) · Philips C3540 · Philips C5-1 · Philips C5-2 (Cartridge) · Philips C5-2 (Explora) · Philips C5-2 (HD3) · Philips C5-2 · Philips C5 · Philips C5040 (21373A) · Philips C5040 (21373B) · Philips C6-3 · Philips C6514 (21372A) · Philips C7-3 · Philips C8-4V · Philips C8-4V (Cannon) · Philips C8-4v (Cartridge) · Philips C8-4v (Explora) · Philips C8-4V · Philips C8-5 (Cartridge) · Philips C8-5 (Explora) · Philips C9-2 · Philips C9-2-PUREWAVE · Philips C9-3IO · Philips C9-3V · Philips C9-4 (HD11) · Philips C9-4 (HD3) · Philips C9-4 (iU22) · Philips C9-4EC · Philips C9-4V · Philips C9-5 ICT · Philips C9-5ec (HD11) · Philips C9-5ec (iU22) · Philips C9-5EC · Philips C9 · Philips CA5-2 (21425A) · Philips CA5-2 · Philips CL15-7 · Philips CLA76 · Philips CT8-4 · Philips D1009 (21224A) Pencil/CW · Philips D1914C-21221A-PENCIL · Philips D1914C-21221B-PENCIL · Philips D1914C-21222A-PENCIL · Philips D1914V-21228A-PENCIL · Philips D1914V-21228B-PENCIL · Philips D1914V-21229A-PENCIL · Philips D2CWC · Philips D2TCD-PEDOFF-PW · Philips D2tcd · Philips D5009V-21231A-PENCIL · Philips D5014V-21223A-PENCIL · Philips D5CWC · Philips E6509 (21336A) · Philips E6514 (21370A) · Philips E7511 (21371A) · Philips eL18-4 · Philips S4-2 · Philips L10-4LAP · Philips L1038 · Philips L1038A (21376A)

Compatible Ultrasound Consoles & Brand Hubs (134)

GE Voluson 730 Pro · GE Voluson 730 Pro V · GE Voluson e · GE Voluson E10 · GE Voluson E6 · GE Voluson E8 · GE Voluson E8 Expert · GE Voluson I · GE Voluson i BT14 · GE Voluson S10 · GE Voluson S10 Expert · GE Voluson S8 · HD 11XE · HD 15 · HD 7 · HD 9 · HD11 · HD15 · HD9 · Hitachi Aloka Arietta 65 · Hitachi Aloka Arietta 850 · Hitachi Aloka Arietta Precision · Hitachi Aloka Arietta Prologue · Hitachi Aloka Arietta V60 · Hitachi Aloka Arietta V70 · Hitachi Aloka F31 · Hitachi Aloka F37 · Hitachi Aloka Hi Vision Ascendus · Hitachi Aloka Prosound 2 · Hitachi Aloka ProSound F75 · Hitachi EUB-2000 · Hitachi EUB-400 · Hitachi EUB-405 · Hitachi EUB-410 · Hitachi EUB-415 · Hitachi EUB-420 · Hitachi EUB-450 · Hitachi EUB-500 · Hitachi EUB-515 · Hitachi EUB-525 · Hitachi EUB-5500 · Hitachi EUB-555 · Hitachi EUB-6000 · Hitachi EUB-6500 · Hitachi EUB-7000 · Hitachi EUB-7500 · Hitachi EUB-8000 · Hitachi EUB-8500 · Hitachi EUB-900 · Hitachi Hi Vision Avius · Hitachi Hi Vision Preirus · Hitachi Noblus · HP ImagePoint · HP Sonos 1000 · HP Sonos 2000 · HP Sonos 2500 · HP Sonos 4500 · HP Sonos 5500 · HP Sonos 7500 · IE33 · IU22 · LOGIQ 3 · LOGIQ 400 · LOGIQ 5 · LOGIQ 5 Expert · LOGIQ 7 · LOGIQ 7 Dimension · LOGIQ 9 · LOGIQ A5 · LOGIQ e · LOGIQ E9 · LOGIQ E9 with XDClear · LOGIQ F8 · LOGIQ i · LOGIQ P · LOGIQ p5 · LOGIQ P5 Premium · LOGIQ P5dd · LOGIQ P6 · LOGIQ P7 · LOGIQ P8 · LOGIQ P9 · LOGIQ portable · LOGIQ S · LOGIQ S6 · LOGIQ S7 · LOGIQ S8 · LOGIQ series · LOGIQ V2 · LogiqBook · M-Turbo · Medison Accuvix V10 · Medison Accuvix V20 · Medison Accuvix X8 · Medison Accuvix XQ · Medison MySono 201 · Medison SonoAce 128BW · Medison SonoAce Pico · Medison SonoAce SA-3200 · Medison SonoAce SA-5000 · Medison SonoAce SA-5500 · Medison SonoAce SA-600 · Medison SonoAce SA-6000 · Medison SonoAce SA-6000C · Medison SonoAce SA-6000II · Medison SonoAce SA-8000 · Medison SonoAce SA-8000EX · Medison SonoAce SA-8000Live · Medison SonoAce SA-8000SE · Medison SonoAce SA-8800 · Medison SonoAce SA-8800MT · Medison SonoAce SA-9900 · Medison SonoAce X4 · Medison SonoAce X5 · Medison SonoAce X6 · Medison SonoAce X8 · Medison SonoVet SV-1500HD · Mindray Consona N6 · Mindray Consona N8 · Mindray Consona N9 · Mindray DC 3Vet · Mindray DC 60 · Mindray DC N2 · Mindray DC N2Vet · Mindray DC N3Vet · Mindray DC-30 · Mindray DC-40 · Mindray DC-6 · Mindray DC-7 · Mindray DC-70 · Mindray DC-8 · Mindray DC-8 Expert · Mindray DC-80 · Mindray DC-80A

All Clinical Engineering & Diagnostic Guides (76)

How to Repair the Ultrasound Probe? · Where Can I Repair the Ultrasound Probe? · Repair Fukuda FUT · Repair Aloka UST · Repair GE C36 Probe: Ultimate Guide to C… · Who repairs Alpinion L3 · Where can BK 8670 ultrasound probe be re… · Who can repair BK 8819/E10C4/9019 ultras… · When you need to repair Konica L18 · Ultrasound Probe Element Dropout: Diagno… · Degraded Ultrasound Image Quality: Probe · Ultrasound Probe Lens Delamination: Bubb… · Ultrasound Probe Lens Scratches and Memb… · Cracked or Dented Ultrasound Probe Housi… · Ultrasound Probe Strain Relief Damage an… · Ultrasound Probe Cable Damage: Jacket We… · Ultrasound Probe Connector Damage: Pins,… · TEE Probe Repair: Bite Damage, Articulat… · Endocavity Probe Fluid Ingress and Leak … · 3D/4D Probe Mechanical Failures: Motor a… · Repair or Replace an Ultrasound Probe: A… · Reparo de Transdutor TEE: Mordida, Artic… · Dano no Cabo do Transdutor de Ultrassom:… · Dano no Conector do Transdutor de Ultras… · Falha de Elementos em Transdutor de Ultr… · Carcaça de Transdutor de Ultrassom Trinc… · Qualidade de Imagem de Ultrassom Degrada… · Delaminação da Lente do Transdutor de Ul… · Riscos na Lente e Desgaste da Membrana d… · Reparar ou Trocar um Transdutor de Ultra… · Dano e Reparo no Alívio de Tensão do Tra… · Falhas Mecânicas em Transdutores 3D/4D: … · Infiltração de Fluido e Teste de Vazamen… · Fallas Mecánicas en Transductores 3D/4D:… · Infiltración de Fluido en el Transductor… · Reparación de Transductor TEE: Daño por … · Daño del Cable del Transductor de Ultras… · Daño del Conector del Transductor de Ult… · Falla de Elementos en Transductor de Ult… · Carcasa del Transductor de Ultrasonido A… · Calidad de Imagen de Ultrasonido Degrada… · Delaminación de la Lente del Transductor… · Rayones en la Lente y Desgaste de la Mem… · Reparar o Reemplazar un Transductor de U… · Daño y Reparación del Protector de Cable… · Probe Fault or System Fault? A Three · Before You Ship an Ultrasound Probe for … · How to Evaluate a Used or Refurbished Ul… · A One · Transdutor de Ultrassom: Tipos, Estrutur… · Manutenção Preventiva de Transdutores de… · Transdutor Endocavitário: Tipos, Cuidado… · Transdutor Linear: Aplicações, Frequênci… · Transdutor Convexo: Uso, Frequências e F… · Transdutor Setorial (Phased Array): Uso … · Transdutor Convexo GE: Modelos, Compatib… · Transdutor Mindray: Modelos, Compatibili… · Transdutor de Ultrassom Philips: Modelos… · Calibração de Ultrassom: O Que É, O Que … · Falha do Transdutor ou do Aparelho? Árvo… · Antes de Enviar um Transdutor para Repar… · Como Avaliar um Transdutor Usado ou Reco… · Rotina Mensal de Verificação de Transdut… · Transdutor Siemens: Modelos, Compatibili… · Transdutor Esaote: Modelos, Compatibilid… · Transdutor Toshiba: Modelos, Compatibili… · Ultrasound Probe Model Cross · Ultrasound Probe Lens & Crystal Arti… · Tabela Cruzada de Modelos de Transdutor … · Tabla Cruzada de Modelos de Transductor … · Autoavaliação de Artefatos e Falhas em T… · Autoevaluación de Artefactos y Fallas en… · Ultrasound Probe Not Recognized & Re… · Transdutor de Ultrassom Não Reconhecido … · Ultrasound Biopsy Needle Guides: Reusabl… · Transductor de Ultrasonido No Reconocido…