The Alpinion C1-6 is a common candidate for repair rather than replacement. This page lists its typical failure modes, self-checks you can run, and the details a repair lab needs from you.
| Manufacturer | Alpinion |
|---|---|
| Model | C1-6 |
| Probe type | Convex array |
| Typical applications | Abdominal, OB/GYN |
| Compatible systems | E-Cube 7 |
The C1-6 typically pairs with E-Cube 7; the same fault can present differently between console revisions, which is why the host system is asked for with the quote details.
Curved arrays like the C1-6 see heavy abdominal duty, and the curved lens face takes the brunt of it – delamination and membrane wear typically show up before element failures do.
Common failure modes for the C1-6
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Depth performance clearly below what this probe used to deliver | Ageing elements or lens delamination reducing acoustic output | Case by case |
| Vertical shadowing or missing scan lines on screen | Element dropout, or a conductor break inside the cable | Case by case |
| Image cuts out when the cable is flexed | Broken internal conductors in the cable | Case by case |
| Visible nicks, grooves or staining on the lens surface | Wear from use, or chemical attack from disinfectants | Usually repairable |
| Outer cable sheath showing cracks, kinks or staining | Surface jacket wear – the conductors inside are often still good | Usually repairable |
| Grainy, low-detail image even after presets are checked | A worn lens surface or elements beginning to degrade | Case by case |
| Split seams or impact marks on the probe body | A drop or knock; housings are routinely replaced or resealed | Usually repairable |
Check these before you ship the probe
- Compare penetration and uniformity against a known-good probe on the same preset.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Run the system’s built-in probe test or element check if your console provides one.
- Flex the cable gently along its length while imaging – intermittent dropout points to broken conductors.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
Repair, exchange or replace — which applies to your C1-6
For the C1-6, think in terms of damage depth: surface and cable-level faults are usually repairable at a fraction of replacement cost; acoustic stack failures are evaluated case by case; extensive element loss generally means replacement. Send the symptoms and photos first – that determines the path before anything ships.
What we need to quote your C1-6 repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the C1-6 probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.