Use this reference to troubleshoot the Esaote SC3123: identify the fault from what you see, judge whether it is typically repairable, and prepare the details needed for a quote.
| Manufacturer | Esaote |
|---|---|
| Model | SC3123 |
| Probe type | Micro-Convex array |
| Typical applications | Small Parts, Vascular |
| Compatible systems | MyLab 60MyLab |
You will usually see the SC3123 running on MyLab 60MyLab. Mention the console model with the fault description – connector and revision differences matter.
Small-footprint probes such as the SC3123 concentrate handling stress on a compact lens and thin cable – mechanical wear usually precedes any element trouble.
Common failure modes for the SC3123
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Bent, broken or corroded connector pins | Connector damage from forced insertion or drops | Usually repairable |
| Fuzzy or noisy image that cleaning does not fix | Lens wear or early element degradation | Case by case |
| Visible nicks, grooves or staining on the lens surface | Mechanical wear or disinfectant damage to the lens | Usually repairable |
| Cable jacket cracked, wrinkled or discolored | Surface jacket wear – the conductors inside are often still good | Usually repairable |
| Noticeably weaker penetration than the paired probe | Reduced acoustic output from element ageing or a delaminating lens | Case by case |
| Air pockets or peeling visible on the acoustic lens | Lens delamination from the acoustic stack | Usually repairable |
| Vertical shadowing or missing scan lines on screen | Element dropout, or a conductor break inside the cable | Case by case |
Check these before you ship the probe
- Flex the cable gently along its length while imaging – intermittent dropout points to broken conductors.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Run the system’s built-in probe test or element check if your console provides one.
Repair, exchange or replace — which applies to your SC3123
For the SC3123, think in terms of damage depth: surface and cable-level faults are usually repairable at a fraction of replacement cost; acoustic stack failures are evaluated case by case; extensive element loss generally means replacement. Send the symptoms and photos first – that determines the path before anything ships.
What we need to quote your SC3123 repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the SC3123 probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.





