Use this reference to troubleshoot the Philips D2CWC: identify the fault from what you see, judge whether it is typically repairable, and prepare the details needed for a quote.
| Manufacturer | Philips |
|---|---|
| Model | D2CWC |
| Probe type | Pencil/CW array |
| Frequency reference | 2 MHz |
| Typical applications | Cardiac, Vascular |
| Compatible systems | Philips Affiniti 50, Philips Affiniti 70, Philips Epiq 5, Philips Epiq 7, Philips iE33, Philips iU22 |
| Source record | The Ultrasound Source product record |
You will usually see the D2CWC running on Philips Affiniti 50, Philips Affiniti 70, Philips Epiq 5, Philips Epiq 7, Philips iE33 and other listed systems. Mention the console model with the fault description – connector and revision differences matter.
For CW pencil probes such as the D2CWC, the short parts list means diagnosis is quick: cable continuity, connector condition and crystal output cover most failures.
Common failure modes for the D2CWC
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Cracked or dented housing | A drop or knock; housings are routinely replaced or resealed | Usually repairable |
| Fuzzy or noisy image that cleaning does not fix | Lens wear or early element degradation | Case by case |
| Large areas of the array not firing | Widespread element or acoustic stack failure | Usually not economical |
| Dark bands or dropout lines in the image | Element dropout, or a conductor break inside the cable | Case by case |
| Strain relief separated, split or stiff | Flex fatigue where the cable meets the probe or connector | Usually repairable |
| Air pockets or peeling visible on the acoustic lens | Lens delamination from the acoustic stack | Usually repairable |
| Cable jacket cracked, wrinkled or discolored | Jacket wear; conductors may still be intact underneath | Usually repairable |
Check these before you ship the probe
- Compare penetration and uniformity against a known-good probe on the same preset.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Run the system’s built-in probe test or element check if your console provides one.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
Repair, exchange or replace — which applies to your D2CWC
Whether a D2CWC is worth repairing depends on the damage scope: localized lens, cable, strain relief and connector faults are routine repairs, while widespread element loss usually is not economical. An exchange unit can bridge the gap when downtime matters, and replacement becomes rational when repair cost approaches replacement value. A written assessment settles this per probe – condition, not model, decides it.
What we need to quote your D2CWC repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the D2CWC probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.






Reviews
There are no reviews yet.