Before writing off a damaged BK Medical 8567-S, compare its symptoms against the failure modes below. Many lens, cable and connector faults on this probe class are economically repairable.
| Manufacturer | BK Medical |
|---|---|
| Model | 8567-S |
| Probe type | Endocavity array |
| Compatible systems | BK Medical Falcon 2101 |
In service the 8567-S is most often found on BK Medical Falcon 2101. Console-side settings and module revisions still vary, so the exact system build is part of any assessment.
For endocavity heads such as the 8567-S, reprocessing chemistry is the main enemy – seal integrity and lens condition decide service life more than the electronics do.
Common failure modes for the 8567-S
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Intermittent signal that changes as the cable moves | Broken internal conductors in the cable | Case by case |
| Large areas of the array not firing | Extensive element loss across the acoustic stack | Usually not economical |
| Split seams or impact marks on the probe body | Impact damage; housing can usually be replaced or resealed | Usually repairable |
| Fuzzy or noisy image that cleaning does not fix | A worn lens surface or elements beginning to degrade | Case by case |
| Latch or locking hardware damaged on the connector shell | Broken locking mechanics on the connector housing | Usually repairable |
| Cuts, gouges or discoloration on the lens membrane | Wear from use, or chemical attack from disinfectants | Usually repairable |
| Pin damage or corrosion visible in the connector | Connector damage from forced insertion or drops | Usually repairable |
| Fluid found inside the probe head after a leak test | Seal failure allowing disinfectant ingress | Case by case |
| Failed electrical leakage test | Compromised sealing on an endocavity probe | Case by case |
Check these before you ship the probe
- Run the system’s built-in probe test or element check if your console provides one.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Have the probe leak-tested before further use – fluid ingress worsens quickly and creates a patient-safety issue.
Repair, exchange or replace — which applies to your 8567-S
Whether a 8567-S is worth repairing depends on the damage scope: localized lens, cable, strain relief and connector faults are routine repairs, while widespread element loss usually is not economical. An exchange unit can bridge the gap when downtime matters, and replacement becomes rational when repair cost approaches replacement value. A written assessment settles this per probe – condition, not model, decides it.
What we need to quote your 8567-S repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the 8567-S probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.





