BK Medical 8804 Ultrasound Probe Repair
Fault diagnosis and repair options for the BK Medical 8804: failure modes, self-checks and what an assessment needs.
Table of Contents
PRODUCT DETAILS
Supply Ability
Supply Ability 5000 Unit/Units per Year
Packaging & Delivery
| Packaging Details | Common Box |
| Port | Hongkong |
| Picture Example: | ![]() ![]() |
DESCRIPTION
The BK Medical 8804 is a common candidate for repair rather than replacement. This page lists its typical failure modes, self-checks you can run, and the details a repair lab needs from you.
| Manufacturer | BK Medical |
|---|---|
| Model | 8804 |
| Probe type | Linear array |
| Compatible systems | BK Medical Falcon 2101, BK Medical Hawk 2102, BK Medical Pro Focus 2202 |
In service the 8804 is most often found on BK Medical Falcon 2101, BK Medical Hawk 2102, BK Medical Pro Focus 2202. Console-side settings and module revisions still vary, so the exact system build is part of any assessment.
On linear probes such as the 8804, superficial imaging workloads put steady stress on the lens membrane and strain relief; the element electronics tend to outlast the mechanical parts.
Common failure modes for the 8804
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Pin damage or corrosion visible in the connector | Mechanical damage from forcing or dropping the connector | Usually repairable |
| Strain relief separated, split or stiff | Flex fatigue where the cable meets the probe or connector | Usually repairable |
| Outer cable sheath showing cracks, kinks or staining | Surface jacket wear – the conductors inside are often still good | Usually repairable |
| Visible nicks, grooves or staining on the lens surface | Wear from use, or chemical attack from disinfectants | Usually repairable |
| Whole regions of the image blank across the array width | Widespread element or acoustic stack failure | Usually not economical |
| Vertical shadowing or missing scan lines on screen | Element dropout, or a conductor break inside the cable | Case by case |
| Latch or locking hardware damaged on the connector shell | Broken locking mechanics on the connector housing | Usually repairable |
Check these before you ship the probe
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Flex the cable gently along its length while imaging – intermittent dropout points to broken conductors.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Run the system’s built-in probe test or element check if your console provides one.
Repair, exchange or replace — which applies to your 8804
Repair suits the 8804 when damage is mechanical or confined – lens membrane, strain relief, cable jacket, connector hardware. Once element dropout spreads across large sections of the array, repair cost climbs toward replacement value and stops making sense. If uptime is critical, ask about exchange options while your probe is assessed.
What we need to quote your 8804 repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the 8804 probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.
Company Profile



Sample Room

Customer Feedbacks




Exhibition




Production Process





Packing & Shipping




FAQ
Q: What is the MOQ?
A: 1pc.
Q: How about your delivery time?
A: 1~15 days depends on your order quantity.
Q: What is your terms of delivery?
A: DHL,TNT, EMS,UPS,FedEx,also can use customer’s freight forwarder.
Q: What is your terms of payment?
A: Western union, Alibaba Trade Assurance, T/T in advance.
Q: Do you test all your goods before delivery?
A: Yes, we have 100% tested our goods before delivery.
Q: What is your Service & Return policy?
A: If the probe is not compatible with your machine or there are any quality problem within 7 days after receiving the goods, please feel free to contact us. We will send a replacement probe or refund the payment to you.







