Before writing off a damaged BK Medical E10C4 (9019), compare its symptoms against the failure modes below. Many lens, cable and connector faults on this probe class are economically repairable.
| Manufacturer | BK Medical |
|---|---|
| Model | E10C4 (9019) |
| Probe type | Endocavity array |
| Compatible systems | BK Medical bk3000 |
In service the E10C4 (9019) is most often found on BK Medical bk3000. Console-side settings and module revisions still vary, so the exact system build is part of any assessment.
For endocavity heads such as the E10C4 (9019), reprocessing chemistry is the main enemy – seal integrity and lens condition decide service life more than the electronics do.
Common failure modes for the E10C4 (9019)
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Whole regions of the image blank across the array width | Extensive element loss across the acoustic stack | Usually not economical |
| Bent, broken or corroded connector pins | Connector damage from forced insertion or drops | Usually repairable |
| Latch or locking hardware damaged on the connector shell | Broken locking mechanics on the connector housing | Usually repairable |
| Noticeably weaker penetration than the paired probe | Ageing elements or lens delamination reducing acoustic output | Case by case |
| Grainy, low-detail image even after presets are checked | Lens wear or early element degradation | Case by case |
| Dark bands or dropout lines in the image | Failed transducer elements or a broken conductor in the cable | Case by case |
| Cracked or dented housing | A drop or knock; housings are routinely replaced or resealed | Usually repairable |
| Fluid found inside the probe head after a leak test | Seal failure allowing disinfectant ingress | Case by case |
| Failed electrical leakage test | Compromised sealing on an endocavity probe | Case by case |
Check these before you ship the probe
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Run the system’s built-in probe test or element check if your console provides one.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Have the probe leak-tested before further use – fluid ingress worsens quickly and creates a patient-safety issue.
Repair, exchange or replace — which applies to your E10C4 (9019)
Repair suits the E10C4 (9019) when damage is mechanical or confined – lens membrane, strain relief, cable jacket, connector hardware. Once element dropout spreads across large sections of the array, repair cost climbs toward replacement value and stops making sense. If uptime is critical, ask about exchange options while your probe is assessed.
What we need to quote your E10C4 (9019) repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific E10C4 (9019) can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.





