Before writing off a damaged Hitachi Aloka C22P, compare its symptoms against the failure modes below. Many lens, cable and connector faults on this probe class are economically repairable.
| Manufacturer | Hitachi Aloka |
|---|---|
| Model | C22P |
| Probe type | Convex array |
| Compatible systems | Hitachi Aloka Arietta 850, Hitachi Aloka Arietta Precision, Hitachi Aloka Arietta Prologue, Hitachi Aloka Arietta V60, Hitachi Aloka Arietta V70, Hitachi Noblus |
You will usually see the C22P running on Hitachi Aloka Arietta 850, Hitachi Aloka Arietta Precision, Hitachi Aloka Arietta Prologue, Hitachi Aloka Arietta V60, Hitachi Aloka Arietta V70 and other listed systems. Mention the console model with the fault description – connector and revision differences matter.
On convex probes such as the C22P, the curved acoustic window wears unevenly under pressure-heavy scanning; lens condition is the first thing to check when image quality drops.
Common failure modes for the C22P
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Noticeably weaker penetration than the paired probe | Reduced acoustic output from element ageing or a delaminating lens | Case by case |
| Pin damage or corrosion visible in the connector | Mechanical damage from forcing or dropping the connector | Usually repairable |
| Vertical shadowing or missing scan lines on screen | Failed transducer elements or a broken conductor in the cable | Case by case |
| Connector lock broken or hardware missing | Mechanical damage to the connector shell | Usually repairable |
| Strain relief separated, split or stiff | Flex fatigue where the cable meets the probe or connector | Usually repairable |
| Whole regions of the image blank across the array width | Widespread element or acoustic stack failure | Usually not economical |
| Grainy, low-detail image even after presets are checked | A worn lens surface or elements beginning to degrade | Case by case |
Check these before you ship the probe
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Run the system’s built-in probe test or element check if your console provides one.
Repair, exchange or replace — which applies to your C22P
Repair suits the C22P when damage is mechanical or confined – lens membrane, strain relief, cable jacket, connector hardware. Once element dropout spreads across large sections of the array, repair cost climbs toward replacement value and stops making sense. If uptime is critical, ask about exchange options while your probe is assessed.
What we need to quote your C22P repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific C22P can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.





