If your Hitachi Aloka UST-987-7.5 is producing degraded images, failing probe tests or showing physical damage, this page helps you judge what is likely wrong and whether a repair makes sense before you commit to a replacement.
| Manufacturer | Hitachi Aloka |
|---|---|
| Model | UST-987-7.5 |
| Probe type | Convex array |
| Compatible systems | Aloka SSD-1700, Aloka SSD-2000, Aloka SSD-4000 |
In service the UST-987-7.5 is most often found on Aloka SSD-1700, Aloka SSD-2000, Aloka SSD-4000. Console-side settings and module revisions still vary, so the exact system build is part of any assessment.
Curved arrays like the UST-987-7.5 see heavy abdominal duty, and the curved lens face takes the brunt of it – delamination and membrane wear typically show up before element failures do.
Common failure modes for the UST-987-7.5
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Fuzzy or noisy image that cleaning does not fix | Lens wear or early element degradation | Case by case |
| Strain relief pulling away from the cable or gone rigid | Cable flex fatigue at the probe or connector end | Usually repairable |
| Depth performance clearly below what this probe used to deliver | Reduced acoustic output from element ageing or a delaminating lens | Case by case |
| Split seams or impact marks on the probe body | A drop or knock; housings are routinely replaced or resealed | Usually repairable |
| Visible nicks, grooves or staining on the lens surface | Wear from use, or chemical attack from disinfectants | Usually repairable |
| Air pockets or peeling visible on the acoustic lens | Lens delamination from the acoustic stack | Usually repairable |
| Latch or locking hardware damaged on the connector shell | Mechanical damage to the connector shell | Usually repairable |
Check these before you ship the probe
- Compare penetration and uniformity against a known-good probe on the same preset.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
Repair, exchange or replace — which applies to your UST-987-7.5
Repair suits the UST-987-7.5 when damage is mechanical or confined – lens membrane, strain relief, cable jacket, connector hardware. Once element dropout spreads across large sections of the array, repair cost climbs toward replacement value and stops making sense. If uptime is critical, ask about exchange options while your probe is assessed.
What we need to quote your UST-987-7.5 repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific UST-987-7.5 can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.