A failing Philips mC7-2 does not always mean buying a new probe. Use the fault table below to match what you are seeing on screen to a likely cause, then check what a repair assessment would need.
| Manufacturer | Philips |
|---|---|
| Model | mC7-2 |
| Probe type | Convex array |
| Compatible systems | Philips Epiq 7 |
You will usually see the mC7-2 running on Philips Epiq 7. Mention the console model with the fault description – connector and revision differences matter.
Curved arrays like the mC7-2 see heavy abdominal duty, and the curved lens face takes the brunt of it – delamination and membrane wear typically show up before element failures do.
Common failure modes for the mC7-2
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Bent, broken or corroded connector pins | Mechanical damage from forcing or dropping the connector | Usually repairable |
| Fuzzy or noisy image that cleaning does not fix | Lens wear or early element degradation | Case by case |
| Vertical shadowing or missing scan lines on screen | Element dropout, or a conductor break inside the cable | Case by case |
| Cracked or dented housing | Impact damage; housing can usually be replaced or resealed | Usually repairable |
| Air pockets or peeling visible on the acoustic lens | Lens delamination from the acoustic stack | Usually repairable |
| Strain relief separated, split or stiff | Cable flex fatigue at the probe or connector end | Usually repairable |
| Whole regions of the image blank across the array width | Extensive element loss across the acoustic stack | Usually not economical |
Check these before you ship the probe
- Run the system’s built-in probe test or element check if your console provides one.
- Flex the cable gently along its length while imaging – intermittent dropout points to broken conductors.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
Repair, exchange or replace — which applies to your mC7-2
For the mC7-2, think in terms of damage depth: surface and cable-level faults are usually repairable at a fraction of replacement cost; acoustic stack failures are evaluated case by case; extensive element loss generally means replacement. Send the symptoms and photos first – that determines the path before anything ships.
What we need to quote your mC7-2 repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the mC7-2 probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.