Before writing off a damaged Samsung LV3-14A, compare its symptoms against the failure modes below. Many lens, cable and connector faults on this probe class are economically repairable.
| Manufacturer | Samsung |
|---|---|
| Model | LV3-14A |
| Probe type | Linear array |
| Compatible systems | WS80A |
In service the LV3-14A is most often found on WS80A. Console-side settings and module revisions still vary, so the exact system build is part of any assessment.
On linear probes such as the LV3-14A, superficial imaging workloads put steady stress on the lens membrane and strain relief; the element electronics tend to outlast the mechanical parts.
Common failure modes for the LV3-14A
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Image cuts out when the cable is flexed | Broken internal conductors in the cable | Case by case |
| Connector lock broken or hardware missing | Mechanical damage to the connector shell | Usually repairable |
| Fuzzy or noisy image that cleaning does not fix | Lens wear or early element degradation | Case by case |
| Air pockets or peeling visible on the acoustic lens | Lens delamination from the acoustic stack | Usually repairable |
| Whole regions of the image blank across the array width | Widespread element or acoustic stack failure | Usually not economical |
| Split seams or impact marks on the probe body | Impact damage; housing can usually be replaced or resealed | Usually repairable |
| Strain relief separated, split or stiff | Flex fatigue where the cable meets the probe or connector | Usually repairable |
Check these before you ship the probe
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Run the system’s built-in probe test or element check if your console provides one.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
Repair, exchange or replace — which applies to your LV3-14A
Whether a LV3-14A is worth repairing depends on the damage scope: localized lens, cable, strain relief and connector faults are routine repairs, while widespread element loss usually is not economical. An exchange unit can bridge the gap when downtime matters, and replacement becomes rational when repair cost approaches replacement value. A written assessment settles this per probe – condition, not model, decides it.
What we need to quote your LV3-14A repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Product imagery on this page is representative of the LV3-14A probe family and is not a photograph of your specific unit. This page does not claim OEM authorization or universal compatibility. Repair feasibility, turnaround and commercial terms are confirmed in writing after assessment of the actual probe.





