If your Sonosite TEExi is producing degraded images, failing probe tests or showing physical damage, this page helps you judge what is likely wrong and whether a repair makes sense before you commit to a replacement.
| Manufacturer | Sonosite |
|---|---|
| Model | TEExi |
| Probe type | TEE array |
| Compatible systems | Sonosite Edge II, SonoSite M-Turbo |
You will usually see the TEExi running on Sonosite Edge II, SonoSite M-Turbo. Mention the console model with the fault description – connector and revision differences matter.
TEE probes like the TEExi combine articulation mechanics, a long insertion tube and bite exposure – three failure paths surface probes never face, and the reason TEE repairs are assessed with extra care.
Common failure modes for the TEExi
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Strain relief separated, split or stiff | Cable flex fatigue at the probe or connector end | Usually repairable |
| Bubbles or lifting at the lens face | The lens separating from the acoustic stack underneath | Usually repairable |
| Noticeably weaker penetration than the paired probe | Ageing elements or lens delamination reducing acoustic output | Case by case |
| Cuts, gouges or discoloration on the lens membrane | Mechanical wear or disinfectant damage to the lens | Usually repairable |
| Bent, broken or corroded connector pins | Mechanical damage from forcing or dropping the connector | Usually repairable |
| Outer cable sheath showing cracks, kinks or staining | Jacket wear; conductors may still be intact underneath | Usually repairable |
| Whole regions of the image blank across the array width | Widespread element or acoustic stack failure | Usually not economical |
| Bite marks or punctures on the insertion tube | Patient bite damage to the flexible shaft | Case by case |
| Tip articulation stiff, loose or unresponsive | Worn or broken articulation cables | Case by case |
| Failed leak test on a TEE probe | Breach in the insertion tube or tip sealing | Case by case |
Check these before you ship the probe
- Run the system’s built-in probe test or element check if your console provides one.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Do not continue clinical use after suspected bite damage; note the articulation behaviour (stiff, loose, unresponsive) for the assessment.
Repair, exchange or replace — which applies to your TEExi
Three paths apply to a failing TEExi: repair (typical for lens, housing, cable and connector damage), exchange (fastest route back to scanning) or replacement (when the acoustic stack is too far gone). The deciding factors are how much of the array still fires, whether the housing is intact and what downtime costs you. Each probe is judged on its actual condition.
What we need to quote your TEExi repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific TEExi can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.





