A failing Toshiba PVU-770ST does not always mean buying a new probe. Use the fault table below to match what you are seeing on screen to a likely cause, then check what a repair assessment would need.
| Manufacturer | Toshiba |
|---|---|
| Model | PVU-770ST |
| Probe type | Endocavity array |
| Compatible systems | Canon Xario 100G , Canon Xario 200G |
You will usually see the PVU-770ST running on Canon Xario 100G (TUS-X100G), Canon Xario 200G (CUS-X200G). Mention the console model with the fault description – connector and revision differences matter.
For endocavity heads such as the PVU-770ST, reprocessing chemistry is the main enemy – seal integrity and lens condition decide service life more than the electronics do.
Common failure modes for the PVU-770ST
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Air pockets or peeling visible on the acoustic lens | The lens separating from the acoustic stack underneath | Usually repairable |
| Split seams or impact marks on the probe body | Impact damage; housing can usually be replaced or resealed | Usually repairable |
| Intermittent signal that changes as the cable moves | Broken internal conductors in the cable | Case by case |
| Cable jacket cracked, wrinkled or discolored | Surface jacket wear – the conductors inside are often still good | Usually repairable |
| Visible nicks, grooves or staining on the lens surface | Wear from use, or chemical attack from disinfectants | Usually repairable |
| Strain relief separated, split or stiff | Flex fatigue where the cable meets the probe or connector | Usually repairable |
| Large areas of the array not firing | Widespread element or acoustic stack failure | Usually not economical |
| Fluid found inside the probe head after a leak test | Seal failure allowing disinfectant ingress | Case by case |
| Failed electrical leakage test | Compromised sealing on an endocavity probe | Case by case |
Check these before you ship the probe
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Flex the cable gently along its length while imaging – intermittent dropout points to broken conductors.
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Run the system’s built-in probe test or element check if your console provides one.
- Have the probe leak-tested before further use – fluid ingress worsens quickly and creates a patient-safety issue.
Repair, exchange or replace — which applies to your PVU-770ST
For the PVU-770ST, think in terms of damage depth: surface and cable-level faults are usually repairable at a fraction of replacement cost; acoustic stack failures are evaluated case by case; extensive element loss generally means replacement. Send the symptoms and photos first – that determines the path before anything ships.
What we need to quote your PVU-770ST repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific PVU-770ST can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.





