A failing Hitachi Aloka UST-9104-5 does not always mean buying a new probe. Use the fault table below to match what you are seeing on screen to a likely cause, then check what a repair assessment would need.
| Manufacturer | Hitachi Aloka |
|---|---|
| Model | UST-9104-5 |
| Probe type | Convex array |
| Compatible systems | Aloka SSD-1000, Aloka SSD-1400, Aloka SSD-1700, Aloka SSD-3500, Aloka SSD-900 |
The UST-9104-5 typically pairs with Aloka SSD-1000, Aloka SSD-1400, Aloka SSD-1700, Aloka SSD-3500, Aloka SSD-900; the same fault can present differently between console revisions, which is why the host system is asked for with the quote details.
Curved arrays like the UST-9104-5 see heavy abdominal duty, and the curved lens face takes the brunt of it – delamination and membrane wear typically show up before element failures do.
Common failure modes for the UST-9104-5
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Depth performance clearly below what this probe used to deliver | Reduced acoustic output from element ageing or a delaminating lens | Case by case |
| Strain relief pulling away from the cable or gone rigid | Cable flex fatigue at the probe or connector end | Usually repairable |
| Image cuts out when the cable is flexed | Broken internal conductors in the cable | Case by case |
| Visible nicks, grooves or staining on the lens surface | Wear from use, or chemical attack from disinfectants | Usually repairable |
| Pin damage or corrosion visible in the connector | Mechanical damage from forcing or dropping the connector | Usually repairable |
| Split seams or impact marks on the probe body | A drop or knock; housings are routinely replaced or resealed | Usually repairable |
| Fuzzy or noisy image that cleaning does not fix | A worn lens surface or elements beginning to degrade | Case by case |
Check these before you ship the probe
- Photograph the complete manufacturer label – the exact revision matters for parts compatibility.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
Repair, exchange or replace — which applies to your UST-9104-5
Repair suits the UST-9104-5 when damage is mechanical or confined – lens membrane, strain relief, cable jacket, connector hardware. Once element dropout spreads across large sections of the array, repair cost climbs toward replacement value and stops making sense. If uptime is critical, ask about exchange options while your probe is assessed.
What we need to quote your UST-9104-5 repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific UST-9104-5 can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.





