A failing Sonosite TEEXp does not always mean buying a new probe. Use the fault table below to match what you are seeing on screen to a likely cause, then check what a repair assessment would need.
| Manufacturer | Sonosite |
|---|---|
| Model | TEEXp |
| Probe type | TEE array |
| Compatible systems | Sonosite X-Porte |
The TEEXp typically pairs with Sonosite X-Porte; the same fault can present differently between console revisions, which is why the host system is asked for with the quote details.
On TEE heads such as the TEEXp, articulation cables, tube integrity and bite damage dominate the fault picture; electrical leak testing is non-negotiable after any suspected breach.
Common failure modes for the TEEXp
Match what you are seeing against the table below. The right-hand column reflects how faults of this kind typically assess – the actual verdict always depends on the individual probe.
| What you see | Likely cause | Typically repairable? |
|---|---|---|
| Pin damage or corrosion visible in the connector | Connector damage from forced insertion or drops | Usually repairable |
| Image cuts out when the cable is flexed | One or more conductors broken inside the cable bundle | Case by case |
| Vertical shadowing or missing scan lines on screen | Failed transducer elements or a broken conductor in the cable | Case by case |
| Fuzzy or noisy image that cleaning does not fix | Lens wear or early element degradation | Case by case |
| Whole regions of the image blank across the array width | Widespread element or acoustic stack failure | Usually not economical |
| Bubbles or lifting at the lens face | The lens separating from the acoustic stack underneath | Usually repairable |
| Split seams or impact marks on the probe body | Impact damage; housing can usually be replaced or resealed | Usually repairable |
| Bite marks or punctures on the insertion tube | Patient bite damage to the flexible shaft | Case by case |
| Tip articulation stiff, loose or unresponsive | Worn or broken articulation cables | Case by case |
| Failed leak test on a TEE probe | Breach in the insertion tube or tip sealing | Case by case |
Check these before you ship the probe
- Run the system’s built-in probe test or element check if your console provides one.
- Compare penetration and uniformity against a known-good probe on the same preset.
- Examine the connector for bent or corroded pins, cracks and a working locking mechanism.
- Check the strain reliefs at both ends for splits, stiffness or separation from the cable.
- Inspect the lens face under good light for cuts, bubbles, lifting edges or discoloration.
- Do not continue clinical use after suspected bite damage; note the articulation behaviour (stiff, loose, unresponsive) for the assessment.
Repair, exchange or replace — which applies to your TEEXp
Repair suits the TEEXp when damage is mechanical or confined – lens membrane, strain relief, cable jacket, connector hardware. Once element dropout spreads across large sections of the array, repair cost climbs toward replacement value and stops making sense. If uptime is critical, ask about exchange options while your probe is assessed.
What we need to quote your TEEXp repair
- A clear photo of the complete manufacturer and model label on the probe
- A description of the fault: what you see on screen and when it started
- The ultrasound system (console) the probe is used with
- Screenshots or photos of the image problem, if the fault is visual
- Photos of any physical damage – lens, housing, cable or connector
Images shown are representative and are not photographs of an individual unit. Nothing on this page claims OEM authorization or guarantees universal fit. Whether a specific TEEXp can be repaired, and on what terms, is confirmed in writing once the probe has been assessed.





