TEE probe repair carries a burden no other transducer does: the probe goes inside a patient, and the thing that keeps the patient safe is a barrier you cannot see. A transesophageal probe is a bending, articulating, repeatedly disinfected instrument, and every one of those three properties is a wear mechanism. Leak integrity is not a quality checkbox on a TEE repair — it is the repair.
Why TEE probes wear out faster than the array does

The acoustic stack in a TEE probe often outlives everything around it. What fails first is the mechanical and chemical envelope:
The bending rubber. The flexible section behind the tip flexes through thousands of articulation cycles and sits in high-level disinfectant between every one of them. Glutaraldehyde, ortho-phthalaldehyde and hydrogen-peroxide systems are all hard on elastomer over time. The rubber stiffens, crazes, then splits — usually along the outside of the most-used bend.
The articulation cables and control mechanism. Stretched control cables give you a tip that will not hold its angle, or that deflects further one way than the other. Left long enough, a frayed cable can jam the tip in a flexed position, which is a genuinely dangerous failure during a procedure.
The insertion tube. Bite damage, kinking and abrasion. Any breach here is a fluid path into the electronics and a patient-safety failure at the same time.
High-level disinfection is a hard requirement, not a preference
Practitioners who handle TEE probes daily are unambiguous about this, and so are we: a TEE probe that has been in a patient must go through full high-level disinfection before it is handled, packed or shipped. That is your responsibility before the probe leaves your department, and it is not something a repair vendor can do retrospectively.
It is worth naming the tension honestly, though. The same HLD cycles that make the probe safe to reuse are also the thing degrading the bending rubber. That is not an argument for disinfecting less — it is an argument for inspecting the flexible section as part of your disinfection routine, because that is when it is already in your hands.
Leak testing: what it does and does not tell you
A leak test pressurises the probe’s interior and watches for pressure loss. It catches breaches. It does not catch a bending rubber that is one flex away from splitting, and it does not catch an articulation cable that has stretched.
So a passed leak test is necessary and not sufficient. Log your leak tests, but inspect the flexible section visually every time as well — you are looking for surface crazing, loss of gloss, and any point where the rubber has gone stiff relative to the sections either side.
Send images before you send the probe
Most repair shops find out whether a TEE probe is repairable after it has crossed a border. We do it the other way round: send us the on-screen images from your own system, a short video of the fault, and the model and part number. We assess it first.
If we judge it not repairable, we tell you not to ship it. That happens rarely — with this class of probe most faults are recoverable — but when it does, you have not paid freight, waited two weeks, or taken the probe out of service for nothing.
A phone call or an email thread is usually enough. What we need to see: the artifact as it appears on your machine, the probe’s behaviour when the cable is flexed, and any visible damage to lens, housing or connector.
What we repair on a TEE probe
Bending rubber replacement, articulation cable replacement and re-tensioning, insertion tube repair or replacement, control-body mechanism work, connector and cable repair, and array work where the acoustic stack has lost elements.
Every TEE probe leaving us is leak tested and electrical-safety tested, and both results go in the report. We do not return a TEE probe on a visual inspection alone.
Turnaround and shipping
About one week in-house for most TEE work. You cover the freight to us; the return shipment is free, including when a probe turns out not to be repairable. Repairs carry six to twelve months of warranty depending on the probe and the work involved, confirmed with your quote — a fault that returns within that period is repaired free. Shipping from Europe? See how the cross-border round trip works.
Documentation
For TEE probes the written report matters more than for any other type, because your department may need it in a service record. It records the fault as found, what was replaced, the leak test result and the electrical safety result. A sample report is here.
Frequently asked questions
The tip will not hold its angle any more. Is that repairable?
Usually, yes. Loss of angle retention is almost always stretched articulation cables or a worn control mechanism, both of which are replaceable parts. Stop using the probe — a tip that will not hold position is a procedural risk, not just an inconvenience.
Our leak test passes but the bending rubber looks cloudy. Do we keep using it?
Cloudiness and loss of gloss are early elastomer degradation. The leak test is telling you the barrier has not failed yet. Plan the replacement rather than waiting for a failed test, because the failure mode here is a breach during a procedure.
Can you work on a probe that has failed a leak test?
Yes — that is one of the more common reasons a TEE probe comes to us. Tell us it failed the leak test when you send images, and disinfect it fully before shipping.